Reprocessing Option 2 - St. Jude Medical MHV-HHR Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 63
NOTE: Higher temperatures or longer sterilization times may be
used; however; increasing the sterilization time or temperature (not
to exceed 138°C/280°F) of the stated cycles may reduce instrument
life.
Temperature
Sterilization Time
Temperature
Sterilization Time

Reprocessing Option 2

Automated Cleaning, Washer-Disinfector Method
The automated washer-disinfector method described below was
validated using STERIS™ Synergy™ washer-disinfector, model
E3023-2.
Prior to placing instruments in the washer-disinfector, remove visible
blood soil under cold tap water with a soft bristle brush. Secure the
instruments in the washer-disinfector and perform the following
programmed cleaning cycle.
Pre-Vac Sterilization
Low Temperature Cycle
121°C/250°F
30 minutes
Gravity Displacement Sterilization
Low Temperature Cycle
121°C/250°F
30 minutes
High Temperature Cycle
132°C/270°F
3 minutes
High Temperature Cycle
132°C/270°F
6 minutes
8