• Extension handle Models EX-05 and EX2000-R
• 35 and 37 mm mitral valve sizer rings, Models 905-35 and 905-37
INDICATIONS
The handles are indicated for use with those products listed in the
Description section of this manual.
CONTRAINDICATIONS
The handles are contraindicated for use with any devices not listed in
the Description section of this manual.
Any sterilization method other than steam is contraindicated.
COMPONENTS
The product package includes one handle model MHV-HHR,
MHV-HHL1, MHV-HHL2, or MHV-HHL3.
STORAGE
The non-sterile handles may be stored indefinitely.
PRECAUTIONS
• Instruments must be cleaned and sterilized prior to each use.
• Do not use cracked, deformed, discolored/rusted, or damaged
instruments.
• Improper cleaning may result in an immunological or toxic
reaction.
• The sterilization temperature must not exceed 138°C (280°F).
3