• Perform a visual inspection of the RPM to assure that all visible contamination has
successfully been removed. If not, repeat the pre-cleaning and cleaning steps.
• Place the RPM in an automatic cleaning container.
• Place in an automated washer/disinfector with the following (default) cycle:
Phase
Pre-Wash
Wash 1
Wash 2
Rinse 1
Thermal Rinse
Pure Water Rinse
Drying
• After the automated wash cycle is complete, blow any excess
water out of the connectors with compressed air.
Settings
Circulation time (Min and Max)
Water type
Temperature
Circulation time (Min and Max)
Injection Range
Water type
Temperature
Circulation time (Min and Max)
Injection Range
Water type
Temperature
Circulation time (Min and Max)
Water type
Temperature
Circulation time (Min and Max)
Water type
Temperature
Circulation time (Min and Max)
Water type
Temperature
Circulation time (Min and Max)
Water type
Temperature
0:15 -15:00
Cold tap water
< 16° C
1:00- 5:00
¼ - 2oz
Hot tap water
≥ 43° C
1:00 - 15:00
¼ - 2oz
Hot tap water
≥ 43° C
00:15 - 10:00
Hot tap water
≥ 43° C
00:15 - 10:00
Hot tap water
≥ 82° C
00:15 (non-recirculation)
Cold tap water
< 16° C
00:00-60:00
NO WATER
66° -115.5° C