Technical assistance must only be performed by qualified Ivory
technicians, who can also provide original Ivory spare parts.
CAUTION!
Do not attempt to disassemble the product or tamper with
any mechanism. This medical product must be used in the
ways illustrated in this manual. The products and systems
specified in the manual for cleaning and maintenance
must also be used. Users are responsible for regular
functioning checks, maintenance and inspection of the
product.
WArNING!
Sterilise the handpiece and all accessories before use.
CAUTION!
Dirty or humid cooling air can cause malfunctioning and
rapid bearing wear. Cooling air must be clean, dry and
uncontaminated, and must conform to ISO 7494-2 standard.
30
Eng
CAUTION!
Only use clean, undamaged burs to prevent dirt from
penetrating into the locking system.
DANGEr!
A bur that is loose, unsuitable or improperly seated can
work free of the head or break. Before use always confirm
that the bur is correctly seated and locked into place.
CAUTION!
Insufficient cooling can lead to overheating of the
preparation and even damage the tooth. Make sure that
the water flow rate is greater than 50NL/min.
WArNING!
Do not use the handpiece to move the patient's cheek
while the device is functioning!