11.2 System Settings

Select your language via the  icon Push  to confirm and to return to the previous menu
Adjust the brightness via the  icon Push  to confirm and to return to the previous menu
Adjust the Volume via the  icon Push  to confirm and to return to the previous menu
Hand-piece Counter Status:
The counter status for the hand-piece that is currently connected, is shown in this display field.
Version info:
Press the version info button to open the window with information about the current software version of
the device.
Restore defaults:
Press the restore default button to reset the factory default settings.
Touch Calibration:
Press the "Touch Calibration" button to open the screen to carry out the touch calibration.
This can be done to improve the touch input if it is not sufficiently accurate.
First press the + symbol in the top left corner. A + symbol will appears in the lower right corner.
Then precisely press the + symbol in the lower right corner.
Repeat the procedure to complete the touch calibration.

11.3 SD CARD

User-defined settings and the treatment recommendation list are saved on the SD card.
NOTE:
If the SD card is not inserted, the message 'SD card not found' appears when the Clinical Protocols and
Favorites buttons are pressed.
Deactivate the message by pressing the button "OK" and continue.