4 Assembly
4.1
Set the on/off switch to the "Off" position. Turn the quick-disconnect
wrist to the left up to the stop. Carefully insert the electronic control
unit into the Electric Greifer until your hear it click in place (Fig. 5).
4.2
Bring the quick-disconnect wrist to mid-position. Set the on/off
switch to the "On" position. Mount the coaxial bushing and make
sure that it's correctly positioned (Fig. 6).
5 Adjusting the Electrical Power Shut-Off
5.1
Position the quick-disconnect wrist so that the dial becomes visible.
5.2
Adjust the dial 1 so to achieve a maximal grip force of 160N.
5.3
With a new Electric Greifer, the shut-off current has been set.
For further information on the operational modes of the Digital Twin con-
trol, please refer to the 647H382 Instructions for Use of the Ottobock
System Electric Greifer with Digital Twin control.
6 Liability
Otto Bock Healthcare Products GmbH, hereafter referred to as manufactu-
rer, assumes liability only if the user complies with the processing, operating
and maintenance instructions as well as the service intervals. The manu-
facturer explicitly states that this product may only be used in combination
with components that were authorised by the manufacturer (see instruc-
tions for use and catalogues). The manufacturer does not assume liability
for damage caused by component combinations which it did not authorise.
The product may only be opened and repaired by authorised Otto Bock
technicians.
Ottobock | 7