Figure 35
√ Utility
D2 Lamp
W Lamp
Printe
3. Set printer
On the function selection display, move the cursor with ▼/▲ (2.9) to select Printer and press ENTER (2.7) to enter
the printer setting display(Fig. 36)
Figure 36
√ Utility
D2 Lamp
on/off
W Lamp
on/off
Printe
Move the cursor with ▼/▲ (2.9) to select Printer style and press ENTER (2.7):
- If printer is outside the equipment select Table style.
- If printer is inside the equipment select Panel style.
Whenever changing the roll of paper, move the cursor with ▼/▲ (2.9) to select Scroll paper and press ENTER
(2.7). Repeat the operation until the new paper spreads out.
Move the cursor with ▼/▲ (2.9) to select Reset and press ENTER (2.7); the printer will be reset.
4. Set time and datec
On the function selection display, move the cursor with ▼/▲ (2.9) to select Time and date and press ENTER (2.7)
to enter the time and date setting display (Fig. 37).
Input the data of hour and date by using the keypad (2.10). Use ▼/▲ (2.9) to move the cursor.
Figure 37
√ Utility
Date & time
Dark current
Reset WL.
5. Dark Current (0%T) (Fig. 38)
Zuzi recommends to carry out this function whenever changing the environmental conditions such as tempera-
ture, voltage or environmental light. For this purpose:
On the function selection display, move the cursor with ▼/▲ (2.9) to select Dark current and press ENTER (2.7).
After a few seconds, the equipment will automatically return to the function selection display and will be ready
to operate.
Figure 38
√ Utility
Date & time
Dark current
Reset WL.
6. Wavelength calibration (Reset WL) (Fig. 39)
Zuzi recommends calibrating the wavelength whenever suspecting of it. For this purpose:
On the function selection display, move the cursor with ▼/▲ (2.9) to select Reset WL and press ENTER (2.7).
After a few minutes, the equipment will automatically return to the function selection display and will be ready to ope-
rate.
Version 2 Nov-10
ENTER
on/off
on/off
RETURN
√ Printer
ENTER
Printer style
Scroll paper
Reset
RETURN
ENTER
RETURN
Auto return
Instruction manual 54251050
√ D2 Lamp
ON
√
OFF
√ Printer
Date & time
Time: 09: 18: 00
Date: 2005-01-01
Week: Monday
...Gatting Dark Current...
Testing paper
Page. 31