Arc2 ON Time
Arc2 OFF Time
Rearc Time
Est. Mode
MFD-L
MFD-R
Minimum Loss
Core Step coefficient
Core curve coefficient
MFD Mismatch
4.3.
Creating splice mode
4.2.1 Default Splice Mode
when the machine is produced, there are three Splice Modes
(Auto, Calibrate, Normal) and 12 splice parameter files (No. 1~12)
©Copyright Abacanto Digital SA.
the total of the Arc 1 time and Arc2 time exceeds 30
seconds, always adjust the function [Arc2 ON Time] and
[Arc2 OFF Time] to weaken the arc power. A continuous
arc over 30 seconds, without weakening the arc power,
may damage the arc discharge unit.
During Arc2 discharge, arc power can be pulsed by
turning the arc on and off. This sets the amount of time
that Arc2 is ON.
Sets Arc2 OFF Time during Arc2. When the Arc2
discharge is intermittent, re-arc discharge is also
intermittent. When continuous re-arc discharge is
necessary, set this parameter to "OFF".
Sets Rearc Time.
Selects splice loss estimation mode to "OFF", "CORE",
or "CLAD". When MM fibers are spliced, select
"CLAD".
Sets MFD of the left and right fibers. Both MFD-L & R
are taken into account for estimating splice loss.
This amount is added to the estimated splice loss
originally calculated. When splicing specialty or
dissimilar fibers, a high actual splice loss may occur even
with optimized arc conditions. To make the actual splice
loss concur with the estimated splice loss, set the
minimum value of estimate to the minimum optimized
actual splice loss.
Determines how Core step, Core curve and MFD
mismatch influences splice loss estimate. If Est. Mode is
set "OFF" or "CLAD", these Core step, Core curve and
MFD mismatch are automatically set "OFF". If the
estimated splice on certain fiber combinations needs
adjustment, Core step, Core curve and MFD mismatch are
used. These are advanced splicer functions and should
be discussed with your representative before changing.