LERBAKKEN 10, 8410 ROENDE DENMARK
9116
For safe installation of 9116B the following must be observed. The module shall only be installed by
qualified personnel who are familiar with the national and international laws, directives and
standards that apply to this area.
Year of manufacture can be taken from the first two digits in the serial number.
4501
For Installation in Zone 2 / Division 2 the following must be observed.
The 4501 programming module is to be used solely with PR electronics modules. It is important
that the module is undamaged and has not been altered or modified in any way. Only 4501
modules free of dust and moisture shall be installed.
Hazardous Classified Location
Class I / II / III, Division 1, Group A,B,C,D,E,F,G
Class I Zone 0 / 1 / 2 Group IIC, IIB, IIA or
Class I Zone 20 / 21
Simple Apparatus or
Intrinsic safe apparatus
with entity parameters:
Vmax (Ui) ≥ Vt (Uo)
Imax (Ii) ≥ It (Io)
Pi ≥ Pt (Po)
Ca ≥ C
+ Ci
cable
La ≥ L
+ Li
cable
+
+
T
+
-
-
Loop
V
4- wire
3- wire
2- wire
Potm . 2
Potm . 3
Potm . 1
RTD / LinR
Revision date:
2010-02-12
FM Installation drawing
-20 ≤Ta ≤ +60ºC
4501
54
53
52
+
-
51
mA
44
43
42
-
+
41
9116
TC
Power Rail
Version Revision
V5 R0
9116 - Product Version 9116-002
Unclassified Location
or
Hazardous Classified Location
Class I, Division 2 Group A,B,C,D T4
Class I, Zone 2, Group IIC, IIB, IIA T4
Status relay, terminal (33,34)
Non hazardous area installation
Voltage max:
Power max:
Current max:
Zone 2 installation:
Voltage max:
Power max:
Current max:
34
33
Relay output, terminal (13,14)
32
Non hazardous area installation
Voltage max:
31
Power max:
Current max:
Zone 2 installation
14
Voltage max:
Power max:
13
Current max:
12
(terminal 11,12,13,14)
11
(terminal 31,32,33,34)
(terminal 91,92,93,94,95)
U
253 V max. 400 Hz
m:
Prepared by:
PB
9116QF01
125 VAC / 110 VDC
62.5 VA / 32 W
0.5 A AC / 0.3 ADC
32 VAC/ 32 VDC
16 VA / 32 W
0.5 A AC / 1 A DC
250 VAC / 30 VDC
500 VA / 60 W
2 A AC / 2 ADC
32 V AC / 30 VDC
64 VA / 60 W
2 A AC / 2 ADC