Remove Transport Securing Device; Approximate Taring - Seca 725 Manual De Instrucciones

Ocultar thumbs Ver también para 725:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
2. Remove transport securing
device
Remove the O-ring which secures the slid-
ing weights.
Retain this for further transport operations.
3. Secure sliding weight bar
Secure the sliding weight bar by tilting the
locking device downwards.
8
4. Place tray on scale
Place the tray centrally on the scale.
The locking levers must engage under the
housing.
5. Attach tray
Attach the tray to the scale by pressing
down the locking levers.
6. Release lock

7. Approximate taring

For the scale to display the correct weight,
it must first be set to 0 kg/lbs.
To do so, push both sliding weights on the
sliding weight bar to 0. If the hand and
counter-hand do not line up, the 0 value
must be set using the tare weight. To do so,
undo the knurled nut (see Fig. 8) and push
the tare weight onto the rod until the hand
and counter-hand line up.