86768G Quark_Translated.2.qxd
Owner's Guide
CAUTION: Only use
Coltène/Whaledent's Light Lens
(Cat. No. C7974). Do not use other
manufacturer's light lens with this
product.
CAUTION: Remove the
handpiece from the charger base
before attempting to replace the
light lens. Do not activate the
handpiece.
• Remove the light lens by rotating
the lens counterclockwise until
the lens disengages from the
handpiece.
CAUTION: Do not touch the LED
under the light lens as hand oil
contaminates the LED and can
compromise performance.
VIII. INSPECTING THE
LIGHT LENS
• Inspect the light lens before each
use to insure that restorative
material has not accidentally
cured to the lens. Replace lens
with Coltène/Whaledent Inc. Light
Lens (Cat. No. C7974) if material
has cured to the lens.
CAUTION: Never look directly
into the light emitted from the
light lens or reflected from tooth
or other surfaces.
IX. INSPECTING THE
CHARGER BASE POWER
CORD
• Periodically inspect the charger
base power cord for frayed
areas or broken insulation.
X. CLEANING
• All external components may be
cleaned by wiping with a soft
cloth moistened with isopropyl
alcohol.
8/14/06
9:23 AM
Page 5
IMPORTANT: Do not use petroleum-
based solvents, iodophors, or
phenolic-based products on the
handpiece or charger base.
(Iodophors and phenolics can stain
the surface of the unit and damage
the plastic housing.)
XI. INFECTION CONTROL
• To comply with infection control
procedures the light shield is to be
placed on the handpiece first, then
slide the barrier sleeve over the
handpiece and shield. A new
barrier sleeve should be used for
each patient.
XII. SPECIFICATIONS
A. Electrical Safety
®
LED is a
Coltolux
tested and listed device
under UL and
CSA standards.
This product fulfills the
requirements concerning
electromagnetic compatibility in
accordance with MDD directive
93/42/EEC.
B. Dimensions (approximate)
Handpiece
Length:
22.1 cm / 8.7"
Probe Dia:
1.22 cm / 0.48"
Grip Dia:
2.5 cm / 1.00"
Charger Base
Height:
7.4 cm / 2.9"
Width:
6.4 cm / 2.5"
Depth:
8.9 cm / 3.5"
C. Operating Voltage/Power
Power Supply
Input:
100 – 240 VAC,
50 – 60 Hz
Output: 5 VDC, 6 watts
Class:
II
Handpiece
Operating Voltage: 4.2 – 3.0 VDC
D. Environmental Conditions
®
Coltolux
LED is designed to operate
in ambient temperatures from 50°F –
104°F (10°C – 40°C), relative
humidity 80% at 77°F (25°C).
CAUTION: To maximize patient
comfort, do not operate
handpiece for more than one
minute continuously if the
ambient temperature is above
86°F (30°C).
E. Classifications & Symbols
®
Coltolux
LED is classified as
IEC60601-1 Class II, Type BF, for
protection against electric shock.
IEC Symbol
Publication: 348
Attention: Consult
accompanying documents
IEC Symbol
Publication: 878-02-03
Type BF Equipment
IEC Symbol
Publication: 416
Class II Equipment