Using a fresh, moist disinfectant wipe, such as
CaviWipes
of the LCD screen of the Handheld Monitor.
Discard the wipe. Repeat with a fresh wipe for
a minimum of thirty (30) seconds. Discard
the wipe.
Using a fresh, moist CaviWipes
wipe the grip section of the Handheld Monitor.
Discard the wipe. Repeat with a fresh wipe for
a minimum of thirty (30) seconds. Discard
the wipe.
Using a fresh, moist CaviWipes
wipe the seams of the Handheld Monitor for
one (1) minute.
Using a fresh, moist CaviWipes
wipe all surfaces of the Handheld Monitor for a
minimum of ten (10) seconds. Discard the wipe.
Allow the wiped and wetted Handheld Monitor
to stand for a minimum of three (3) minutes.
At the end of the three (3) minutes, use a sterile
swab saturated with 70% Isopropyl alcohol (IPA)
to thoroughly swab the seams around the edge
of the button panels for a minimum of fifteen
(15) seconds.
Use a sterile 70% IPA wipe, or a sterile,
lint-free wipe, saturated with sterile 70% IPA,
to thoroughly wipe the entire Handheld Monitor
for a minimum of thirty (30) seconds.
Allow the device to air dry. Once dry, visually
inspect the Handheld Monitor for any residual
soil. Repeat cleaning/disinfection instructions if
visible soil is present.
Place the cleaned/disinfected device in an
appropriate dry storage area.
8.3 Low Level Combined Cleaning/Disinfecting
Instructions
After each patient use, power off and unplug all
electric connections.
®
, thoroughly wipe the front and back
®
, thoroughly
®
, thoroughly
®
, thoroughly