NewLife Family
Specifications
Flow Rates*
Dimensions
Weight
Sound Pressure
Level
Power
Consumption
O2 Concentration
Output Pressure
Operating
Environment*
Altitude
Storage
Environment
Warranty
Maintenance
Schedule
Max Tubing
* Based on an atmospheric pressure range of 700 hPa to 1060 hPa at 70° F (21° C)
** At altitudes below sea level and higher than 8,000 ft (2438 m) above sea level, flow meter accuracy may be affected
up to 13%.
The expected service life of this device is a minimum of five years.
See technical manual (PN MN240-1) for sound power level.
10 - ENG
PN MN239-C4 C | User Manual
NewLife Elite
1–5 LPM
±10% of indicated setting, or 200 mL,
whichever is greater**
28.5 x 15.7 x 14.5 in
(72.4 x 40.0 x 36.8 cm)
54 lbs (24.5 kg)
53 dB(A) at flow rates of 1 to 5 LPM
350 watts–1-5 lpm model
Two-prong polarized plug
Double insulated cabinet
North American Models:
120 VAC, 60 Hz, 4.0 amps
Export Models:
230 VAC, 50 Hz, 2.0 amps
230 VAC, 60 Hz, 2.0 amps
1-3 lpm: 95.5% - 92.0%
4 lpm: 92% ± 3%
5 lpm: 90% ± 3%
7.3-8.8 psi (50.3-60.7 kPA)
41° to 104° F (5° to 40° C)
15–95% Humidity
-1250 to 10,000 ft (-381 to 3048 m)
(tested to 700 – 1060 hPa)
-25° C to 70° C (-13° F to 158° F),
15–95% Humidity (non-condensing)
3 Years
Felt Filter (230 V only) – 1 Year
Replacement
Intake Filter – Clean Weekly
50 ft (15.2 m)
NewLife Intensity 10
2–10 LPM
±10% of indicated setting, or 200 mL,
whichever is greater**
27.5 x 16.5 x 14.5 in
(69.9 x 41.9 x 36.8 cm)
58 lbs (26.3 kg)
58 dB(A) at flow rates of 2 to 10 LPM
600 watts–2-10 LPM model
Two-prong polarized plug
Double insulated cabinet
North American Models:
120 VAC, 60 Hz, 6.0 amps
Export Models:
230 VAC, 50 Hz, 3.0 amps
2–9 LPM: 92%+3.5%/-3%, 10 LPM:
90%+5.5%/-3%
20 psig (138 kPA)
41° F to 104° F (5° C to 40° C), 15–95%
humidity
-1250 to 10,000 ft (-381 to 3048 m)
(tested to 700 – 1060 hPa)
-25° C to 70° C (-13° F to 158° F), 15-95%
Humidity (non-condensing)
3 Years
Felt Filter – 1 Year Replacement, Intake
Filter – Clean Weekly
200 ft (61 m)