NewLife Family
WARNING: THIS DEVICE SUPPLIES HIGH-
CONCENTRATION OXYGEN THAT PROMOTES
RAPID BURNING. DO NOT ALLOW SMOKING
OR OPEN FLAMES WITHIN THE SAME ROOM
OF (1) THIS DEVICE, OR (2) ANY OXYGEN-
CARRYING ACCESSORY. FAILURE TO OBSERVE
THIS WARNING CAN RESULT IN SEVERE FIRE,
PROPERTY DAMAGE AND / OR CAUSE PHYSICAL
INJURY OR DEATH.
WARNING: DO NOT USE YOUR OXYGEN
CONCENTRATOR IN THE PRESENCE OF
FLAMMABLE GASES. THIS CAN RESULT IN RAPID
BURNING CAUSING PROPERTY DAMAGE, BODILY
INJURIES OR DEATH.
WARNING: DO NOT LEAVE A NASAL CANNULA
ON CLOTHING, BED COVERINGS OR CHAIR
CUSHIONS. IF THE UNIT IS TURNED ON BUT NOT
IN USE, THE OXYGEN WILL MAKE THE MATERIAL
FLAMMABLE. SET THE I/0 POWER SWITCH TO
THE 0 (OFF) POSITION WHEN THE OXYGEN
CONCENTRATOR IS NOT IN USE.
WARNING: USE NO OIL, GREASE, OR
PETROLEUM-BASED OR OTHER FLAMMABLE
PRODUCTS WITH THE OXYGEN-CARRYING
ACCESSORIES OR THE OXYGEN CONCENTRATOR.
ONLY WATER BASED, OXYGEN COMPATIBLE
LOTIONS OR SALVES SHOULD BE USED.
OXYGEN ACCELERATES THE COMBUSTION OF
FLAMMABLE SUBSTANCES.
WARNING: THE USE OF SOME OXYGEN
ADMINISTRATION ACCESSORIES NOT SPECIFIED
FOR USE WITH THIS OXYGEN CONCENTRATOR
MAY IMPAIR ITS PERFORMANCE.
RECOMMENDED ACCESSORIES ARE
REFERENCED WITHIN THIS MANUAL
CAUTION: Always operate the unit in
an upright position.
12 - ENG
PN MN239-C4 C | User Manual
Proper Setting of Oxygen Flowmeter
To set the proper flow of supplemental oxygen, turn
the flowmeter adjustment knob left or right until the
ball inside the flowmeter centers on the flow line
number prescribed by your physician.
Elite (top) and Intensity (bottom)
To view the flowmeter at the proper angle, note that
the back line and the front numbered line must give
the appearance of just one line.
CAUTION: The Manufacturer
recommends an alternate source of
supplemental oxygen in the event
of a power outage, alarm condition,
or mechanical failure. Consult your
physician or Equipment Provider for
the type of reserve system required.
CAUTION: It is very important to select only the
prescribed level of oxygen. Do not change the
flow selection unless you have been directed to
do so by a licensed clinician.
Normally, you should not need to adjust
the flowmeter on your unit. If you turn the
flowmeter adjustment knob clockwise, you
will decrease and can shut off the flow of
oxygen from your unit. For your convenience,
the flowmeter is marked in ½ LPM increments.
For units equipped with the 2 LPM flowmeter
option, the flowmeter is marked in 1/8 LPM
increments for flow settings up to 2 LPM.
The Oxygen Concentrator may be used during
sleep under the recommendation of a licensed
clinician.