MedComp 14F Split Cath Rg Instrucciones De Uso página 11

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 13
Warning: Do NOT pull distal end of catheter through incision as contamination of
wound may occur.
10.
Apply pressure to proximal tunnel for approximately 10-15 minutes or until
bleeding stops.
11.
Suture incision and apply dressing in a manner to promote optimal healing.
12.
Check catheter for integrity when removed.
Flow vs. Pressure:
Note: Table demonstrates average pressure of arterial and venous lumens during
a simulated dialysis treatment at the respective flow rates indicated on the table.
Fluid used was 55% Saline and 45% Glycerine with a viscosity similar to blood
(3 to 4 centipose).
Priming Volume:
14F Split Cath Rg
Length
24cm (1
24cm (2
40cm
40cm (2
Size
ml/min
14F x 24cm
Venous
Arterial
14F x 40cm
Venous
Arterial
cut)
st
cut)
nd
(1
cut)
st
cut)
nd
Blood Flow Rate
300
400
ml/min
mmHg
mmHg
64
100
-71
-112
mmHg
mmHg
90
140
-90
-143
Venous
1.9cc
1.8cc
2.5cc
2.4cc
-9-
Arterial
1.8cc
1.7cc
2.4cc
2.3cc

Publicidad