Model RAD
11. ERROR CODES
During the initial power-on testing or during operation it is possible the scale may show an error message. The meaning of the error
messages is described below.
If an error message is shown repeat the procedure that caused the message, turning the balance on, calibration or other functions. If
the error message still is shown then contact your dealer for further support.
ERROR CODE
Err 4
Err 5
Err 6
Err 7
Err 8
Err 9
Top loading scale
DESCRIPTION
Initial zero is greater than
allowed (4% of maximum ca-
pacity) when power is turned
on or when the
key is
pressed.
Keyboard error.
A/D count is not correct when
turning the scale on.
Can't set current weight as
100%.
Calibrate weight error
Unstable, can't return to zero.
POSSIBLE CAUSES
Weight on the pan when turning the scale on.
Excessive weight on the pan when zeroing the scale.
Improper calibration of the scale. Damaged load cell.
Damaged electronics.
Improper operation of the scale.
Platform not installed.
Load cell damaged.
Electronics damaged.
This message only show when you press
Use incorrect weight.
This message only show during turn on power, please keep the
scale stable or warm the scale for some minutes.
User's Guide
key in zero point.
29