9PIN D CONNECTER:
Scale
Pin 2
Pin 3
Pin 3
3. CONTINUOUS & STABLE OUTPUT PROTOCOL
"CONT":
S
T
,
-HEADER 1-
4. COMMUNICATION PROTOCOL
The XTA & XTA-P scale has the following protocols, accessible from the R232 menu:
Toledo
NCI - ECR
NCI - General
TEC
Easy
POS
10. ERROR CODES
Error Message
Description
Maximum load exceeded
----
Zero setting error
Err 4
Key board error
Err 5
A/D value out of range
Err 6
Percentage Error
Err 7
Calibration weight error
Err 8
Unstable Reading
Err 9
Gravity error
Err 15
Tare out of range
Err 17
Over range
-- ol ---
Underload
-- lo ---
fAil H
Calibration Error
fAil l
fAil
bA lo
Battery low
lo bA
16
Computer
Pin 3
Pin 2
Pin 5
G
S
:
-
-HEADER 2-
Solution
Unload or reduce weight
Zero setting range exceeded due to switching on. (4%max) Make sure platform
empty(4%max) Make sure platform empty.
Check the keys and connecter
Make sure platform empty and check the pan is installed proper. Check the load cell
connectors.
Please check input data, must be > 0.5d
Check the test weights for calibration or linear calibration
Check any air variation, vibration, RF noise and touching some where. Check the
load cell and connecters
Check the gravity settings. Gravity range must be 0.9xx ~
Minus weight or overload. Remove the load and restart
Remove the load. Re calibrate
Minus weight, check the platform and restart or calibrate.
Check the test weights & Re calibrate
Re charge battery, check the voltages.
+/-
- WEIGHT DATA -
USER MANUAL XTA
EN
g
k
CR
LF
-WEIGHT UNIT-
-TERMINATOR-