Error Message; Weight Conversion - Baxtran BAP Serie Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Several seconds later, once it take stability, the display shows the weight value of third calibration mass (full
capacity) Place the displayed weigh on the pan.
Several seconds later, the buzzer tweets. Then the calibration procedure is finished. Switch off the balance &
take away all the weights; then switch on again and the balance can be used.
Note:
bility whithout have the properly mass on the pan, it will assume stable weigh as
asked weigh, producing a wrong calibration. You must press ESC key for to start
calibration process again.
When balance is on, the display may show E1 or E2, which can be eliminated by recalibrating the balance.
Before calibrating the balance, it must have been turned on for at least 30 min in the room where the calibration
is to be carried out.

8. ERROR MESSAGE

During operation, the scale may show some error messages.
CODES
E1
E2
E5
OL
Correct your operation and contact the dealer or manufactory for further assistance.

9. WEIGHT CONVERSION

200.00
300.00
-If during the loading of calibration mass process scale arrive to take sta-
CODES
CAUSES
E1
The Span calibration data loses
E2
The initial zero drifts when power on.
POSSIBLE CAUSES
EPROM data loss or MCU is damaged
The zero of the load cell is out of the initial zero range.
The sample weight is not enough
Overload
CONVERSION UNITS TABLE FOR WEIGHTS
1 ct (Metric. Carat)
1 oz (AVOIRDUPOIS OUNCE)
1 GN (GRAIN) (UK)
1 ozt (TROY OUNCE)
1 dwt (PENNYWEIGHT)
1 t (TOLA, India)
USER MANUAL BAP
g
g
=
0.1999694
g
=
28.349523
g
=
0.0647989
g
=
31.103476
g
=
1.555174
g
=
11.663804
g
EN
EN
19

Publicidad

Tabla de contenido
loading

Tabla de contenido