Operation
– Display application peaks
– Delete application peaks
To open the quick menu:
1. Select
in measuring mode.
2. Select and confirm the desired feature.
4.9 Opening information
1. Select
> Info in measuring mode.
The following options are available:
Option
Messages
Device information
Gas statistics
32
Description
The pending warnings and
errors are displayed.
Device information is dis-
played.
The following gas statis-
tics are available:
– Select Shift peak to
display the exposure
peaks for all gasses.
– Select Application
peak to display the
application peaks for
all gasses.
– Select TWA values to
display the available
TWA values for all gas-
ses.
– Select STEL values to
display the available
STEL values for all
gasses.
Option
Intervals
Capture ranges
Battery
4.10 Monitoring
4.10.1 Special features when measuring
with the pump
NOTICE
Magnetic media may be damaged!
The pump adapter and calibration adapter contain a
magnet which may delete data from a magnetic stripe.
► Do not bring magnetic media (e.g. credit cards) into
close proximity to the pump adapter or calibration
adapter.
– Use the dust and water separators when taking
measurements with the pump.
– Following a bump test with aggressive gases (such
as biogas or chlorine), flush the pump with clean air
for several minutes to extend the service life of the
pump.
Instructions for use
Description
The following intervals are
available:
– Select the bump test
interval (Bump test
interval) to display the
days remaining until
the next bump test for
all channels. For
details, select and con-
firm the respective
channel.
– Select Calibration
interval to display the
days remaining until
the next calibration for
all channels. For
details, select and con-
firm the respective
channel.
– Select Life time to dis-
play the remaining life
time.
Capture ranges are dis-
played.
The battery state of
charge is displayed
(large).
|
X-am® 3500