Adjustment Philosophy
SWITCH MAX
SWITCH MIN
MODE
ANALOG MAX
ANALOG MIN
Factory Set / RESET
On delivery, the sensor is programmed to factory settings.
With factory reset all modes are inactive.
The factory reset can be done at any time.
When powered on and after reset:
With a correctly connected supply voltage and after a successful self-test, the Power LED of the sensor is
on and a sequential run of all mode LEDs will occur (approx. 2 seconds).
Start of operation:
With a correctly connected supply voltage and successful self-test, the Power LED of the sensor will turn
on and the last programmed settings will be loaded. Depending on the adjustments programmed and the
container's current filling level, the corresponding mode LEDs will be on or off.
16
Mode 4
Mode 3
Mode 2
Mode 1
SET
6 Seconds
+
MODE
Pulse flashing
(all mode LEDs)
The mode changes according to the shown
infinite loop.
SET
By pressing the MODE key the prompt will
be activated for 5 seconds (slow flashing of
the LED, approx.1 Hz)
During these 5 seconds one has 3 choices:
SET
1. Press MODE key
→ Mode change
2. Press SET key
SET
→ Make adjustment in current Mode
(fast flashing of the LED, approx.
10 Hz)
3. Press no key
→ Quit programming mode automatically
SET
Perform a factory reset:
Press MODE and SET key at the
same time for approx. 6 seconds
until the LEDs light up sequentially.
Note: While holding both buttons, all
RESET
of the LEDs will flash.