Start The Application; Data Transmission; Serial Interface Mode; Interface Parameter - Grundfos DIT-IR Installation And Operating Instructions

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33
After starting the setup.exe file, the installation folder can be
selected. It is recommended to confirm the suggested default
folder by clicking [Next].
After confirmation, the software license agreement appears,
which must be accepted.
Find other languages on the CD under
"[CD drive name]:\Installer\Volume\setup.exe".
To continue, click [Next].
A summary of the programs to install appears. To start the
installation, click [Next].
8
The software will be installed.
Complete the installation according to the menu navigation.
After the installation is completed, a restart is necessary. After the
restart, the software can be used.

5. Start the application

Infrared data from the photometer will be received by DIT-IR at
9600 baud. The baud rate can be changed by the supplied
software. See section 5.3.3 Configuration of the DIT-IR settings.
When the unit is switched on, the last selected interface will be
active.

5.1 Data transmission

Position the activated DIT-IR and the transmitting instrument,
such as a photometer, in line of sight, face to face. See also
fig. 5). Start print command at the transmitting instrument.
The data will be sent to the DIT-IR module.
If the printer is not connected, the incoming data
Caution
will be lost!
The DIT-IR does not support and will not indicate
any printer-related errors (paper jam, paper tray
Note
empty, etc.).

5.2 Serial interface mode

To scroll until the serial interface mode is set, press the [Select]
key. The LED in front of the DB9 connector will flash up to
indicate the serial interface mode. All ASCII printers with serial
interface can be used.
5.2.1 Connection to a serial printer
Connect the serial printer to the DB9 connection of the DIT-IR
module. Use a cable with non-transposed conductors.
Pin assignment of the serial cable:
Pole 2 is connected to pole 2 of the other DB9 connection
Pole 3 is connected to pole 3 of the other DB9 connection
GND (pole 5 to pole 5).

5.2.2 Interface parameter

Set the same baud rate on the photometer (see the photometer
installation and operating instructions), the DIT-IR
(see 5.3.3 Configuration of the DIT-IR settings) and the serial
printer (see manual).