HC151 PHRX
Digital multifunction solenoid metering pump
4.2.2 WORKING MODE – PH
In this mode, the pump does a dose proportional to the value Ph read input. You can make a dosage
acid(when the value of the limit of proportionality L1 is greater than the value of the setpoint) or
alkaline (when the setpoint is greater than the value of the limit of proportionality L1).
In both cases, the L1 frequency is set must be greater than the frequency set by the setpoint.
The menu consists:
ENTER
Press and hold the ESC key at any point in the menu to return to working mode menu (see
4.2)
ADSP9500092
MODE
PH
ENTER
SetP:<S1>
F:<FS1> %
L1PH:<L1>
F:<FL1> %
ENTER
LUNGO
PH value
<N>
ENTER
ENABLE USER
YES/NO
ENTER
MODO FUNC
SAVE? YES/NO
14/04/2015 Rev. 1.0
With the Up and Down setting S1 and FS1 (flow
setpoint) and L1PH and FL1 (flow L1) values. S1
and L1 can be set between 0 ... 14pH in 0.1 ph.
FS1 and FL1 1 to 100 in steps of 1. The ENTER
key moves the cursor to set values
<N> is the input read
Use the Up and Down keys to enable the user to
modify the working mode setttings.
Use the Up and Down keys to save this working mode.
YES –press & hold the ENTER key to program pump.
NO - press & hold the ENTER key to return to the
working mode menu
ENGLISH
46/123