9.
Settings Mode
Pressing for the
fourth log in digit
moves you to the
4 Hedson
"Settings mode".
3 Service
2 Supervisor
1 Operator
0 0 0 0
0 Log out
Normal program running mode
Thanks to the advanced control system in this machine
there are many settings that can be altered.
For safety reason these have been divided into several
"levels", that each require a unique code, that is pass-
word.
To enter the "Settings mode" and from there reach the
"Automatic testing mode" you have to press
+ at the same time as the main switch is turned on.
To enter the required passwords in window 1,2,3 and 4
you press until the first zero starts to flash. Use the
or to set the correct first digit. Then press to edit
the second digit in the code and so on.
The default passwords are indicated below:
Operator
0000
Operator or supervisor
7913
Note : The operator password can be changed.
Please see chapter 9.6.
10
Trisk CureMaster 2 DTL
To adjust a setting press until the value starts to
flash. Then use the or
to adjust to the correct
value. To confirm and save your setting press .
}
9 Long dist lim.
Statistics,
8 Short dist lim.
}
Resetting and
7 Operators PWD
6 (Not used)
Self test
5 Buzzer
(chapter 10)
4 Temp unit
3 Process alarm
2 Run program
1 Select language
9.1
Select Language
Language settings. Available are: Danish, German, Eng-
lish, American-English, Dutch, Dutch-B, Norweigan,
Polish, Portugese, Finnish, French, Swedish, Spanish,
Italian,Turkish, Chineese, Russian, Japanese
9.2
Run programme
High and low settings are set to be the same in factory
settings.
Attention: Each time you change settings between "low
/ high programme", all programme values are returned
to factory settings. Individual adjustments must then be
re-entered in accordance with chapter 7.5 or 7.8.
9.3
Process alarm
Set the alarm to be active or not.
When the target temperature deviates more than 30
from actual a warning is displayed:
"Warning! Process error"
9.4
Temp Unit
Celcius (
C) or Farenheit (
F)
o
o
9.5
Buzzer
The buzzer can be overall muted for all functions.
9.6
Operator PWD
(Only displayed after supervisor log in)
The Customer supervisor can change the code to prevent
unauthorised access to the "Settings and automatic
testing" mode.
9.7
Short distance limit
(Only displayed after supervisor log in)
The "too close" limit in distance measuring can be
adjusted. Factory setting is 55 cm.
9.8
Long distance limit
(Only displayed after supervisor log in)
The "too far" limit in distance measuring can be
adjusted. Factory setting is 65 cm.
# 420878 Rev 1
10. Statistics, Resetting and Self test
To enter the Statistics menu, please log in as described in
Settings mode, chapter 9. By pressing 0 or 1 anywhere in
settings mode you will enter the statistics menu.
Statistics
Reset filter timer?
Reset prog. setting?
Perform selftest?
Test buttons
}
Display test (=pixel test)
Output-Buzzer (=sound test)
Output-IR1 (=lamps cass. 1)
C
o
Output-Fan1
Output-IR2 (=lamps cass. 2)
Output-Fan2
Output-Laser
Input-Pyrometer
Input-Distance sensor
Input-Temperatures
Test Completed
Rough filler
(Normal program running mode)
# 420878 Rev 1
10.1 Statistics and resetting
10.1.1 Statistics
Runtime
The runtime figure shows the accumulated time the IR
lamps have been lit.
Start-ups
Shows the total number of starts for the unit.
10.1.2 Reset filter timer?
Runtime
3:23
Only displayed with supervisor log in
Start-ups
34
When runtime reaches 400 hours you will receive a war-
}
ning saying "Replace filter, reset timer".
This warning is an additional window showing up before
starting any program. The user will have to acknowledge
Yes
this message by pressing .
Then the machine will work as normal.
No
}
Attention: A too dirty filter will shorten the life time of
the IR lamps due to reduced cooling caused by the dirty
filter.
10.1.3 Reset program setting?
All values in pre-defined and user defined programs can
be reset to factory settings.
Press
for 3 seconds to
enter next window
Warning!
When pressing I/Yes all your personal adjustments will
disappear.
Note
The high/low overall program settings in chapter
9.2 will not change by this reset.
Trisk CureMaster 2 DTL
11