Setting damping
Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display re-
acts. You can set damping to suit requirements.
– Press the FUNC key and keep it de-
pressed for 2 seconds.
FIL and the current setting appear in
the display:
0 = slight damping
➞ fast determination of weight
1 = average damping
➞ normal determination of weight
2 = strong damping
➞ slow determination of weight
– Press an arrow key if you want to
change the set value.
– Press the FUNC key again to store the
displayed value. The scale is then back
in normal weighing mode.
The settings are retained even once
the scale is switched off.
6. Storing the scale
The scale can be stored upright to save space.
– Undo the knurled nut on the hinge of
the column.
– Pull out the knurled screw and fold over
the column.
– Re-engage the knurled screw.
– You can now prop the scale up on the
column.
– The scale can now be moved.
The scale can be placed on the bottom rail
for short-term storage. Secure the scale
so that it does not fall over when stored in
the upright position.