Setting damping
Damping is useful for reducing weight determination malfunctions (due to patient mo-
vements, for example). The stronger the damping, the more slowly the weight display
reacts. You can set damping to suit requirements.
– Press the FUNC key and keep it de-
pressed for 2 seconds.
FIL and the current setting appear in
the display:
0 = slight damping
⇒ fast determination of weight
1 = average damping
⇒ normal determination of weight
2 = strong damping
⇒ slow determination of weight
– Press an arrow key if you want to
change the set value.
– Press the FUNC key again to store the
displayed value. The scale is then
back in normal weighing mode.
The settings are retained even once the scale is switched off.
6. Cleaning
Clean the rubber surface and the housing as required using a domestic cleaning agent
or commercially-available disinfectant. Follow the manufacturer's instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like
for cleaning. Such substances can damage the high-quality surfaces.