13. Technical Specifications
Standard Measurement Conditions
• Temperature - 25°C +/-5°C
• Load Range - 500 Ohms - 1 kOhms
• Power Supply - 3.0V DC +/-10%
Output Waveforms
The following are theoretical standard measurement output current across purely resistive loads at maximum intensity
setting. Pulse Width and current measured as shown across 1kOhm loads. These measurements are also valid on a 500
Ohm load, as the Phoenix is a current controlled device. Any load between 500 Ohms and 1 kOhm will not affect the output
measurements. Your output may vary depending on parameter settings.
P1 Disuse Atrophy Program
520 Energy levels, corresponding to 100
Maximum Intensity:
mA intensity and 300 μs pulse width
Zero net DC component
2 x 30 μC
Maximum charge per pulse:
32
P2 Reinforcement Program
520 Energy levels, corresponding to 100
Maximum Intensity:
mA intensity and 300 μs pulse width
Zero net DC component
2 x 30 μC
Maximum charge per pulse:
Phoenix User's Manual