Gima 28102 Guía De Uso página 24

Ocultar thumbs Ver también para 28102:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 54
Surge EN 61000-4-5 ± 0.5kV e± 1kV
Voltage dips, short
outages and
voltage variations
EN 61000-4-11
Network frequency
magnetic field
EN 61000-4-8
Note: U T is the value of the supply voltage
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Mistral aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in
such environment.
Immunity
Level set out by
test
EN 60601-1-2
Conducted
3Vrms 150kHz
immunities
to 80MHz
EN 61000-4-6
(for non-life-
supporting
appliances)
Radiated
10 V/m 80MHz
immunity
to 2.7GHz
EN 61000-4-3
(for non-life-equip-
ment appliances)
ENGLISH
differential mode
5% U T (>95% dip
in U T ) per 0.5 cycle
40% U T (60% dip
in U T ) per 5 cycles
70% U T (30% dip
in U T ) per 25 cycles
<5% U T (>95% dip
in U T ) per 5 s
30 A/m
Level of
conformity
V
=
1
3 V rms
E
=
1
10 V / m
The appliance
The power supply should be
does not alter
that of a typical commercial
its status
premise or hospital.
--
The power supply should be
that of a typical commercial
premise or hospital.
If the user of the Mistral aero-
sol requires the appliance to
work continuously, it is rec-
ommended to use it with an
uninterruptible power supply.
The appliance
The power supply should be
does not alter
that of a typical commercial
its status
premise or hospital.
Electromagnetic Environment -
Guide
Portable and mobile RF communication
devices should be used no closer than
the separation distance from any part
of the Mistral device, including cables,
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distanc-
es
d=
V
3.5
[
/
]
1
d=
E
12
[
/
]
1
d=
E
23
[
/
]
1
P is the maximum rated output power
of the transmitter in Watt (W) according
to the manufacturer of the transmitter
and d is the recommended separation
distance calculated in meters (m).
The field strengths from fixed RF trans-
mitters, as established in an electro-
magnetic survey of the site
be less than the level of conformity of
each frequency range
√P
√P
80MHz to 800MHz
√P
800MHz to 2,7GHz
.
b
24
, could
a

Publicidad

Tabla de contenido
loading

Tabla de contenido