Caution - Welch Allyn CL 100 Surgical Illuminator Manual De Mantenimiento Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 71
4
Please observe the following WARNINGS and CAUTIONS when choosing a location:

Caution

The CAUTION sign calls attention to an operating procedure, or the like, which if not
correctly performed or adhered to, could result in damage or destruction of a part of the
WARNING Keep out of reach of patient.
WARNING Not suitable for use in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
WARNING Disconnect from patient before the discharge of a cardiac
defibrillator.
WARNING ELECTRICAL SHOCK HAZARD. Do not remove top cover.
WARNING Routinely examine the power cord and plug. Do not use if
inspection reveals damage. Replace only with approved hospital grade power
cord and plug with appropriate electrical rating.
WARNING Do not place the exposed distal end of the Micro Link
bundle, or endoscope, in direct contact with flammable materials or human skin.
WARNING The Light Source must never be operated without Micro Link
or fiber bundle attached to either a headlight or other accessory. Failure to
observe these precautions may result in burns to skin, clothing or other material
inadvertently placed in front of the Micro Link
WARNING High energy radiated light guided through endoscopes may give rise
to high temperatures in front of the light outlet and to the tip of the instrument.
WARNING Use care not to point the distal end of the fiber or headlight directly
at the eye when unit is operating. The brilliant light output from the unit can
cause eye injury.
WARNING Applied parts attached to this equipment should be approved 1) to
the Medical Device Directive (93/42/EEC)(Europe), 2) approved by a Nationally
Recognized Testing Laboratory (United States), and 3) by a Standards Council of
Canada accredited laboratory (Canada) to the appropriate medical standards. The
use of non-approved components may compromise safety.
WARNING During life threatening procedures, it is recommended to have a
backup Light Source and an additional CL series Replacement Lamp.
WARNING The CL series lamps are highly pressurized, any damage or
improper handling may cause premature failure or explosion of the lamp.
WARNING The CL series lamps are hot immediately after use, allow the lamp
to cool for five (5) minutes after operation. Temperatures peak in the CL Series
light sources approximately 60 seconds after the Light Source is powered off.
CL100 / CL300 Surgical Illuminator
®
Fiber, fiber
®
Fiber or fiber bundle.
®
Fiber

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cl 300 surgical illuminator

Tabla de contenido