depletion (less than 20.8%).
When an alarm setpoint is reached, the:
Audible alarm sounds
Alarm lights flash
Vibrator activates
Type of alarm displays by alternately flashing the ALARM
icon and the LO or HIGH icon, depending on how the Low
and High alarms were set.
Low Alarm indicates:
NOTE: The LOW alarm (lower of the two %O
ing and will not reset when O
LOW setpoint.
To reset alarm push the TEST button
False oxygen alarms can occur due to barometric pressure
(altitude) changes or extreme changes in ambient temperature.
It is recommended that an oxygen calibration be performed at
the temperature and pressure of use. Be sure that the instru-
ment is in known fresh air before performing a calibration.
Special requirements for ALTAIR PRO O
Oxygen Detector:
2-12
the lower %O
a more urgent condition and the faster alarm
sequence will be indicated
"LO" "ALARM" displays.
Operate this instrument only in ambient
temperatures ranging from 0ºC to 40ºC
(32ºF to 104ºF).
If the instrument is taken into an environment where
the ambient temperature is 15ºC (59ºF) or more,
higher or lower, than its previous location, do not
use the instrument for a period of 40 minutes to
allow the sensor to adjust to the new conditions.
After this 40-minute period, perform a Fresh Air
Set-up (FAS) to re-zero the sensor. Perform the
FAS in fresh air only.