5. CONFIGURE TARGET SETTINGS
Press the Mode button to view target settings.
These settings are locked by default.
TARGET DEW-POINT TEMPERATURE
You can set the AIRVO 2 to three target dew-point temperature
settings:
•
37°C (98.6°F)
•
34°C (93°F) [if compliance at 37°C is a problem]
•
31°C (88°F) [for face masks only].
You may not have access to all settings, if:
•
the unit is in Junior Mode (limited to 34 °C),
•
the unit was initially set up with tighter limits.
The AIRVO 2 will return to its default setting (37°C) after every
disinfection cycle.
Press the Mode button to move on to the next screen.
TARGET FLOW
You can set the AIRVO 2 to flows between 10 L/min and 60 L/min, in
increments of 1 L/min (10-25 L/min) and 5 L/min (25-60 L/min).
You may not have access to all settings, if:
•
the unit is in Junior Mode (limited to 2 - 25 L/min,
in increments of 1 L/min),
•
the unit was initially set up with tighter limits.
The AIRVO 2 will remember its target flow setting when you switch it
off.
To change the target flow setting:
Follow the same sequence of steps as above in "To change the
target dew-point temperature setting".
A – 7
To change the target
dew-point temperature setting:
Hold the Up and Down buttons for 3 seconds to
"unlock" the setting.
The lock will disappear and be replaced by an arrow
showing the minimum and maximum accessible
settings. Press the Up and Down buttons to choose
the new setting.
When you have finished, press the Mode button to
'lock' the setting again.
The lock will reappear.