d. Monitor the capsule outlet. When liquid fl ow is reduced to a trickle, reduce the system pressure
to 0 bar (0 psig).
2.C. Instructions To Use Multiple Scale-Up Capsules In Series Or Parallel Flow Path
3M
Emphaze
AEX Hybrid Purifi er Scale-Up capsules can be used in parallel to add capacity to
the process stream. 3M Emphaze AEX Hybrid Purifi er Scale-Up Capsules can be used in series
fl ow path with 3M
Zeta Plus
3M Emphaze Scale-Up capsules, the 3M Emphaze AEX Hybrid Purifi er capsules should be placed
downstream from the 3M Zeta Plus capsules. The capsules can be stacked on top of each other.
To reduce the risk of Injury from possible end user exposure associated with
degradation products released from the purifi cation media; and To reduce the
risk of product loss associated with a) chemical release from purifi cation media
and b) nullifi cation of Supplier's USP<88> Class VI compliance of fi lter capsule
components.
• The capsule must be subjected to a preconditioning fl ush prior to use.
Figure 1. Series Operation
for Two Scale-Up Capsules
V1
IN
V4
V3
OUT
4. Open both vents V4 and V5 if closed.
5. Open V1 slowly to begin fl uid fl ow.
6. When fl uid exits vent V5, close V5, and simultaneously open V2.
7. When fl uid exits vent V4, close V4 and simultaneously open V3.
8. The fl uid remaining on the upstream side may then be processed through the system:
a. Allow the system pressure to go to 0 bar (0 psig).
b. Close V1 and open V5 and reduce the system pressure to 0 bar (0 psig).
c. Apply 0.3 – 0.7 bar (5 – 10 psi) air or nitrogen gas pressure to V5. Using a non-surging pump or
regulated, low pressure, clean air supply, push process fl uid through the capsule. Do not allow
the inlet pressure to pulse or exceed 1.4 bar (20 psig).
d. Monitor the system outlet. When liquid fl ow is reduced to a trickle from the top capsule, reduce
the system pressure to 0 bar (0 psig) and then close V5 and V2.
e. Open V4 and apply 0.3 – 0.7 bar (5 – 10 psi) air or nitrogen gas pressure.
f. Monitor the system outlet. When liquid fl ow is reduced to a trickle, reduce the system pressure
to 0 bar (0 psig) and close V4.

3. Capsule Use with Pre-use Autoclave Sterilization

3.A. Recommended Pre-use Autoclave Sterilization of Capsules
1. Remove the sanitary connection protective covers and vent cap from the capsule.
Scale-Up Capsules. When using combinations of 3M Zeta Plus and
WARNING
For Series Operation for Two Scale-Up
Capsules:
V5
1. Follow the preconditioning fl ush
V2
2. Confi gure the system as shown in Figure 2.
3. To process fl uids, close valves V1, V2,
-8-
IMPORTANT NOTE: Precondition fl ush
each capsule individually prior to use.
procedure given in II. Single Stage Use
Instructions for each capsule used (See
IMPORTANT NOTE).