Software Revisions; Troubleshooting - Stryker AHTO Manual De Utilización Y Mantenimiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51

Software Revisions

Checking the Software Revision Level
1. Depress the Battery Check, , and
2. The current software revision will scroll in the Flowrate Display.

Troubleshooting

Note If the following solutions do not resolve the noted problem, contact
Stryker Endoscopy for repair (see the "Service and Claims" section,
p. 19).
Problem
System does not power on
when cassette is inserted
Motor does not turn on
when irrigation button is
pressed
Low flow rates or system
does not prime
buttons simultaneously.
Ensure cassette is fully inserted into the console, with both latches
clicking into position.
Ensure the AC power cord is properly connected to a hospital
grade power outlet, and the inlet on the rear console panel. The
batteries may need to be charged.
Ensure the proper operation of all fuses. See the 'Fuse
Replacement' section of this manual for further instructions.
Replace tube set as it may be defective.
Ensure cassette is fully inserted into the console, with both latches
clicking into position.
Ensure the cassette contacts and console contacts are clean and
free of debris and buildup.
Replace tube set as it may be defective.
Ensure the Irrigation Bag is not kinked at the exit nozzle.
Ensure the handpiece is at least 18" below the saline bag.
Toggle through the arrows to a higher flow rate setting.
Solution
13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

250-070-601

Tabla de contenido