Cleaning and Maintenance

WARNING Unplug the unit from the electrical outlet before cleaning.
Caution
Do not sterilize the unit.
1. Wipe the unit with a soft cloth dampened in a mild cleaning solution.
2. Clean the unit with disinfectant if needed.

Fuse Replacement

In the event that fuses need replacement, perform the following steps:
1. Locate the fuse holder next to the AC inlet.
2. Squeeze the tabs to remove the fuse holder cover.
3. Remove the old fuses and replace with new 1A 250V fuses.
4. Reattach the fuse holder cover to complete installation.

Battery Pack Replacement

The battery pack should be replaced every 1.5 - 2 years, and must be replaced only
with the Stryker Battery Pack (P/N 250-070-602).
WARNING Never replace the battery pack in the vicinity of the patient.
Installing a New Battery Pack
Follow steps 1-6 in the "Setting Up the Unit" section (p. 8).

Battery Disposal

Follow the proper disposal procedures for nickel-metal hydride cells. As a
minimum standard, observe the following :
• Contact the local environmental agency for information on recycling
nickel-metal hydride batteries.
• Ensure the batteries are fully discharged prior to disposal.
• Do not incinerate.
• Observe all national, state, and local rules and regulations for disposal
of rechargable batteries.
12