• When installing the product, provide space of approximately 10cm around the product for easy access to the inlet and the Power Cord.
• For safety, install the Unit in a place where the AC power cord can be easily removed.
• Read this Operation Manual before use to fully understand the product functions and file for future reference.
• Do not use iCare to lubricate air scalers (except for NSK Ti-Max S970 series) and Phatelus air motors. The iCare lubricating system is not
designed to lubricate those particular products. Maintain air scalers and Phatelus air motors as described in the operation manual for each
product.
• If blood infiltrates inside a handpiece, iCare may not totally clean the internal handpiece components, and internal blood coagulation may
cause product failure. In this case, NSK recommends to immediately lubricate those handpieces with PANA SPRAY Plus / PANA SPRAY.
• Autoclave handpieces AFTER lubricating with this product.
• Place a collector such as a tray under the Unit.
• After use, turn off the Power Switch and shut off the air supply (Air Compressor).
• Drain water from the Air Compressor and Air Filter at least once a week as required. Moisture from the tank could mix with the Maintenance
Oil and defeat the purpose of this device.
• Remove the Air Tube and the AC power cord from the Unit if the Unit is not to be used for a long time.
• Do not store the Maintenance Oil in contact with halogen, strong acid, alkali, and oxidizing substance or in the same place as these
substances.
• Use only consumable parts such as maintenance oil, and accessories such as an AC power cord specified by NSK. Other than what we
specify may cause product failure due to insufficient lubrication.
• This device is for indoor use only.
• Keep the Unit on a level surface.
• Do not attempt to disassemble the product or tamper with the mechanism except as recommended by NSK in this Operation Manual.
• Do not allow any impact on to the product. Do not drop the product.
• Should the product function abnormally, cease operation immediately and contact your Authorized NSK Dealer.
• Perform regular function and maintenance checks.
• If the product is not used for a long period check it is functioning correctly before use.
• Installation and use of this product requires special precautions regarding EMC according to the EMC information.
• Portable and mobile RF communications equipment can affect Medical Electrical equipment. Do not use RF equipment near the product.
• The use of ACCESSORIES such as cables, other than those specified by the manufacturer, with the exception of cables sold by the
manufacturer of this product as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of
this product.
• This product should not be used adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, this product should
be observed to verify normal operation in the configuration in which it will be used.
• The system may present a possibility of malfunction when used in the presence of an electromagnetic interference wave. Do not install the
system in the vicinity of any device which emits magnetic waves. Turn off the Main Power Switch of the system as an ultrasonic oscillation
device or an electrode knife is located close to the vicinity of use.
• U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
• No special training is required for this device.