Master switch on shaker / mixer won't light
1. Check power cord
2. Check circuit breakers
Item won't shake (master switch lighted))
1. Timer run-down
2. Thermal motor circuit breaker triggered by motor overload
Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.
3. Mechanical parts broken (humming motor noise) or electronic failure (no motor noise).
Model
Motion
Rotation speed
Rotation speed setting
Orbit
Operating mode
Input power
Weight
Dimensions (wxhxd)
Standard equipment
Max. load capacity
Supply voltage
Overheating protection
Protection class (DIN 40050)
Permissible ambient temp.
Permissible relative humidity
Designed for continuous use

DISPOSAL

MALFUNCTION & REMEDY

SPECIFICATIONS

Multi Reax
Vibrating
150 – 2000
(rpm)
Digital
(mm)
3
Timer or continuous
(W)
33
(kg)
9,8
(mm)
270 x 410 x 172
Carousel attachment for 26 test tubes, 10 - 16mm
and max. length of 160 mm,
or Carousel attachment for 12 test tubes, 16 -
32mm and max length of 120 mm
(kg)
1,5
230/240V / 50/60 Hz or 115V / 50/60 Hz
Self-resetting
IP 30
0 – 50
(°C)
(%)
80
yes