Inogen One G2 Manual Del Usuario página 18

Ocultar thumbs Ver también para One G2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 101
CAUTION
The Inogen One® G2 battery acts as a secondary power supply in the event of
a planned or unexpected loss of the AC or DC external power supply. When
operating the Inogen One® G2 from an AC or DC external power supply, a
properly inserted Inogen One® G2 battery should be maintained in the unit.
This procedure will ensure uninterrupted operation and will operate all
alarms and alerts in the event of a loss of the external power supply.
NOTE
The Inogen One® G2 Battery requires an initial full charge of uninterrupted
charging from an empty state using the Inogen One® G2 on the AC power
source. Do not run the Inogen One® G2 on battery power until this initial
charging has been completed. Once the initial charge is completed, the battery
may be used with any amount of charge level.
4. Connect the Universal Power Supply.
Connect the AC Power plug to the Universal Power Supply. Connect the
AC Power Plug to the power source and connect the Power Output Plug to
the Inogen One® G2. The green LED on the Universal Power Supply will be
illuminated and a beep will sound from the concentrator.
CAUTION
Ensure the Universal Power Supply is in a well ventilated location as it relies
on air circulation for heat dissipation. The Universal Power Supply may
become hot during operation. Make sure the Universal Power Supply cools
down before handling.
CAUTION
The Universal Power Supply is not water resistant.
CAUTION
Do not disassemble the Universal Power Supply. This may lead to
component failure and/or safety risk.
CAUTION
Do not place anything in the Universal Power Supply port other than the
supplied wall cord. Avoid the use of electrical extension cords with the Inogen
One® G2. If an extension cord must be used, use an extension cord that has
an Underwriters Laboratory (UL) Mark and a minimum wire thickness of 18
gauge. Do not connect any other devices to the same extension cord.
14

Publicidad

Tabla de contenido
loading

Tabla de contenido