Amprobe ALC-110 Manual Del Usuario página 16

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

LIMIT
Limit selection:
The LIMIT function offers 5 limit values (3.5 mA / 10 mA / 12 mA /
0.25 mA / 0.5 mA) in accordance with safety standards.
1. Continue pressing LIMIT button until desired limit value
is selected. The unit will be looping between 3.5 mA /
10 mA / 12 mA / 0.25 mA / 0.5 mA values.
2. When desired value is selected wait 2 sec to automatically
apply selection. The LIMIT symbol will be displayed on the
LCD to indicate that the function is active.
3. Verify the selection by pressing the LIMIT button one
time. The unit will display the selected value for 2 seconds.
4. To disable the LIMIT function, press the LIMIT button for
>2 sec.
Display when limit is exceeded:
The internal buzzer will continue beeping and the
display will flash and show a warning symbol when the
measurements exceeds the selected limit value.
12

Publicidad

Tabla de contenido
loading

Tabla de contenido