CAUTION:
Before performing this procedure, be sure the unit has cooled to
room temperature. The lamp cartridge and IR filters operate at very high
temperatures.
1. Disconnect the illuminator from the power source.
2. Open the lamp access door and remove the lamp, gently rocking from front to
back. Set lamp aside.
3. Remove I.R. filter using a short Phillips head screw driver to remove one Teflon
holder. Replace I.R. filter with new I.R. filter reinstalling Teflon holder.
4. Install the lamp, being sure it is totally seated, and close the access door.
CAUTION:
Before performing this procedure, be sure the unit has cooled
to room temperature.
1.
Disconnect the illuminator from the power source.
2.
Remove the ten cover screws on the sides and the two on top of the unit.
3.
Open lamp access door and lift the top straight up and off the unit. Set aside.
4.
Remove the lamp cartridge assembly, gently rocking from front to back. Set lamp
cartridge aside.
5.
Disconnect the power supply lamp wires from the lamp base igniter module at the
(+) and (–) terminals on the igniter circuit board.
6.
Remove turret unscrewing a phillips head screw that holds a shoulder stud in center
of turret.
7.
Remove the E-external retaining ring holding the shutter assembly to the front panel.
8.
Slide off the shutter assembly from mounting pin. Set aside.
9.
Place the unit on its side.
10. From the bottom, remove the four screws that hold the lamp base to the bottom
panel.
11. Remove the lamp base from the unit. Remove optic/turret block from base by
removing three flat head phillip drive screws. Place optic/turret block on new base.
12.
Place the new lamp base assembly into the unit with the banana jacks in the heat
shield area. Secure with the four screws to the bottom of the unit.
13. Place the unit back on its feet.
14. Replace the shutter assembly. Protruding set screw in shutter assembly must go
between two pressed in pins on sheet metal cabinet casing. Re-install E-external
retaining ring.
15. Reconnect the lamp power supply wires to the igniter module terminals.
16. Install the lamp cartridge, being sure it is totally seated.
17. Replace the top cover making sure the lamp access door is over the lamp cartridge.
18. Replace the twelve cover screws and secure.
19. Replace Turret and shoulder stud using Phillips head screw driver.
LIT194 CUDA SURGICAL
(English)

I.R. FILTER ASSEMBLY REPLACEMENT

LAMP BASE REPLACEMENT