Extended Commands - Vega VEGAPULS C 22 Instrucciones De Servicio

Ocultar thumbs Ver también para VEGAPULS C 22:
Tabla de contenido

Publicidad

13 Anexo

Extended Commands

Command
Break
aXRPOM!
aXWPOM<value>!
aXRDU!
aXWDU<value>!
aXRTU!
aXWTU<value>!
aXRSR!
aXRSR<value>!
Device-Status
11)
Failure
Code
F013
F017
F025
F036
F040
F080
F105
F260
F261
F264
F265
Maintenance
Code
M500
M501
M504
M505
M507
M508
Value 4 with aD0!, aR0!, aRC0!, value 2 with aD0! behind aV!
11)
46
Response
-
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
Description
No measured value available
Adjusted span too small
Error in the linearization table
No executable software
Error in the electronics
General software error
Measured value is determined
Error in the calibration
Error in the device setting
Installation/setup error
Measurement function
Description
Error in the delivery status
Error in the non-active linearization table
Error on an device interface
No measured value available
Error in the device setting
Non executable Bluetooth software
Description
A data recorder starts a request by transmitting a
break
Read Power Operation Mode
Write Power Operation Mode
Read distance unit
Write distance unit
Read temperature unit
Write temperature unit
Read stage reference
Write stage reference
VEGAPULS C 22 • SDI-12

Publicidad

Tabla de contenido
loading

Tabla de contenido