STANDARD CYCLES
CYCLE B
- 134°C: 4 min. sterilisation plus 14 min. drying
- 134°C B Flash*: 3,5 min. sterilisation plus 8 min. drying
- 121°C: 16 min. sterilisation plus 14 min. drying
CYCLE S
- 121°C: 16 min. sterilisation plus 8 min. drying
- 134°C wrapped: 4 min. sterilisation plus 8 min. drying
- 134°C B unwrapped: 4 min. sterilisation plus 6 min. drying
*
maximum recommended load per version 16: 0.5 Kg wrapped load – 0.2 kg porous load. For version 21: 0.7 Kg
wrapped load – 0.3 kg porous load.
NOTE: For each sterilisation cycle available, the manufacturer must carry out Works Tests in compliance with EN 13060
(par. 7.3.1). The cycles available for which the manufacturer has not carried out Works Tests must undergo Installation
Tests according to EN 13060 par. 7.4.
PRION CYCLE
- Ciclo at 134°C: 18 minutes sterilisation plus 14 minutes drying.
COMPACT FLASH CARD USE
To use the Compact Flash card, comply with the instructions supplied on the CD. To
download the data, use the specific program supplied.
"LAST ADVANCED CYCLE" FUNCTION
The "LAST ADVANCED CYCLE" function is activated automatically for all the programs.
30' after the end of the sterilisation cycle, the heating elements switch off and a second drying
cycle automatically takes place when the temperature inside the chamber reaches 60°C.
ISTERILISATION TIMES INTERVENTION
You can increase the sterilisation times; the operation must be carried out by a specialist
technician.