Descargar Imprimir esta página

Agilent Technologies TwisTorr 304 FS Serie Manual De Istrucciones página 168

Ocultar thumbs Ver también para TwisTorr 304 FS Serie:

Publicidad

17
Technical Information
RS 232/RS 485 communication description
N.
203
204
205
206
300
301
302
320
to
399
400
402
404
500
503
504
168/178
Read/Write
Datatype
R
N
R
N
R
N
R
N
R
N
R
N
R
N
Reserved to Agilent service
R
A
R
A
R
A
Reserved to Agilent service
R/W
N
R/W
L
TwisTorr 304 FS on board Controller User Manual / 87-901-028-01 (C)
Description
Driving frequency in Hz
Pump temperature in °C
Pump status
Error code
Cycle time in minutes (zeroed
by the reset command)
Cycle number (zeroed by the
reset command)
Pump life in hours (zeroed by
the reset command)
CRC EPROM (QE)
CRC Param. (PA)
CRC Parameter structure
RS 485 address
Serial type select
Admitted Values
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description:
see the
following figure
0 to 999999
0 to 9999
0 to 999999
QE8XXXX
(where "XXXX"
are variable)
PA8XXXX
(where "XXXX"
are variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)

Publicidad

loading

Este manual también es adecuado para:

X3507-64002X3507-64003