EXTRAORDINARY MAINTENANCE
Extraordinary maintenance includes the repair of any accidental breakage
and the replacement of worn or malfunctioning parts.
To guarantee correct operation of Hurrimix
EN
alginate bowls, XR0250804 dental stone bowls) must be replaced when the
bowl cap no longer clicks when closing.
All electrical and mechanical interventions must only be
performed by a Zhermack-authorised maintenance technician.
• Any modification that alters equipment features from a safety
and risk-prevention point of view can only be carried out by
the Manufacturer who will then certify conformity of the
equipment with safety standards. Therefore any modification
or maintenance intervention not considered in this manual is
to be considered prohibited.
• Any modification, tampering or repair performed by
unauthorised staff along with the use of non-original spare
parts relieve the Manufacturer from any liability. If the
necessary maintenance interventions are not considered in
this Use and Maintenance Manual, contact the Authorised
After-Sales Centre.
• The manufacturer provides the Authorised Support Centre
with circuit diagrams, list of components, and other important
information.
8.4
RE-START AFTER LONG PERIOD OF INACTIVITY
If the equipment is re-started after a particularly long period of
inactivity, operate as indicated in Chapter 6 regarding commissioning.
If the equipment functions anomalously, act as described in Chapter 9
regarding "Diagnostics". if the problem should persist or is not considered,
contact the Authorised After-Sales Centre immediately.
8.5
hOW TO REqUEST AFTER-SALES ASSISTANCE
Choose one of the following options to request after-sales assistance:
1. contact the Authorised After-Sales Centre;
2. telephone the Zhermack After-Sales Service at +39 0425 597611 and
proceed as follows: give personal data (e.g. address and telephone
number); state the information given on the equipment identification
plate (see paragraph 3.3); explain the problem with the equipment clearly;
3. fax all of the information stated in the previous point and the correctly
filled-in identification form (see Attachment 10.2) to the following n° +39
0425 53596.
ChAPTER 9: DIAGNOSTICS
9.1
GENERAL WARNINGS
If any of the anomalies described below should occur, solve the
problem, where allowed, with respect to all instructions
contained in this manual. If the problem persists, contact the
Authorised After-Sales Centre.
9.2
TROUBLEShOOTING
FAULT
When pressing the START button, the mixer does not
work.
CAUSE
1. The plug is not correctly inserted.
2. The main switch was not turned on.
3. The ON/OFF button has not been pressed.
4. The lid was left open.
5. None of the above.
REMEDY
1. Connect the plug properly to the mains.
2. Turn on the switch placed on the rear of the mixer.
3. Press the On button placed at the left of the display.
4. Close the door properly.
5. Contact the Authorised After Sales Centre.
FAULT
Clear noises or vibrations when mixing.
CAUSE
1. The mixing container is too full.
2. The mixing container is empty.
3. Powder and water leak during mixing..
REMEDY
1. The total content should never exceed 100 g (powder +
water).
2. Never activate the device when the container is empty.
3. Always ensure that the mixing container is closed. Ensure
that the cover is not damaged. Do not exceed the maximum
quantities.
The reference figures are present on pages 3, 4, 5 and 6 of this manual and in the product Quick Guide.
28
, the mixing bowls (XR0250802
2
9.3

ERROR MESSAGES

Error messages viewed on display:
error 1: it was attempted to start a mixing cycle with the door open.
error 2: it was attempted to start or restart a mixing cycle without first
having opened the door to remove the previous mixture.
error 3:
motor malfunction (contact technical assistance).
error 4: mains voltage too low.
error 5:
mains voltage too high.
error 6:
circuit board malfunction (contact technical assistance).
error 7:
brake malfunction (contact technical assistance).
error 8:
door lock malfunction (contact technical assistance).
ChAPTER 10: ATTAChMENTS
10.1
EC DECLARATION OF CONFORMITY
The Manufacturer:
Zhermack S.p.A.
Via Bovazecchino, 100
45021 Badia Polesine • RO • Italy
Tel. +39 0425 597611 • Fax +39 0425 53596
Person authorised to compile the technical file:
Zhermack S.p.A.
Via Bovazecchino, 100
45021 Badia Polesine • RO • Italy
Tel. +39 0425 597611 • Fax +39 0425 53596
declares under its sole responsibility that the product:
Hurrimix
2
230 Vac 50/60 Hz
is in conformity with Directive 2011/65 EU of the European Parliament
and of the Council of 8 June 2011 on the restriction of the use of certain
hazardous substances in electrical and electronic equipment.
It is also in conformity with the laws implementing the following directives:
Directive 93/42/EC amended by 2007/47/EC class I medical device (Annex
VII).
According to the following harmonised standards:
EN 60601-1
EN 60601-1-2
The product must be used in compliance with the documentation provided
by the Manufacturer.
Please note that the manufacturer's liability for damages caused by a
defective product expires 10 years from the date on which the manufacturer
released such product on the market and that such liability is transferred to
the user in accordance with Directive 85/374/EEC, subsequent amendments
and the laws implementing it.
Badia Polesine, 31/12/2016
Paolo Ambrosini
General Manager
hurrimix
2

Publicidad

Tabla de contenido
loading

Productos relacionados para Zhermack Hurrimix 2 C305140

Este manual también es adecuado para:

Hurrimix 2 c305141

Tabla de contenido