4.6 Drying
After each cleaning and disinfection, place all product compon-
ents on a dry, clean and absorbent surface and let them dry
completely.
4.7 Storage
Store this product as described below:
• Wrap all individual components in a clean, lint-free cloth
(e.g., a tea towel).
• Store this product in a dry, dust-free place.
5
TROUBLESHOOTING
Fault
No aerosol is
coming out of
the nebuliser.
Possible cause
The nebuliser nozzle in-
sert is blocked.
The compressed air sup-
ply tube is not connected
correctly to the com-
pressed air connection
on the compressor or the
nebuliser lower section.
The tube attached to the
compressed air connec-
tion on the compressor is
leaking.
– 24 –
Remedy
Clean the nebuliser.
Check that the tubing
connectors are connec-
ted firmly to the com-
pressor and the
nebuliser.
Replace the defective
tube or the entire
PARI SINUS tubing
system.