Using a standard thermal disinfector for baby
bottles (not a microwave oven)
CAUTION
Risk of infection due to inadequate disinfection
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Make sure that the disinfector is clean and operating prop-
erly before every disinfection process.
• Allow the disinfection to continue until the disinfector
switches off automatically or the minimum disinfection
time stated in the instructions for use of the disinfector has
elapsed. Do not switch the device off prematurely.
Use a thermal disinfector with a runtime of at least 6 minutes.
Regarding the performance of the disinfection, the duration of
the disinfection procedure and the quantity of water required,
follow the instructions for use of the disinfector you are using.
4.4 Care of the PARI SINUS tubing system
Dry the PARI SINUS tubing system after every use:
• Insert the matching tubing connector on the first of the two
tubes in the compressed air connector on the compressor.
• Switch the compressor on.
• Allow air to blow through the tube until the air has removed
all moisture from the tube.
• Switch the compressor off.
• Disconnect the tube from the compressor, and attach the
other tube to the compressed air connection.
• Repeat the process as described for the first tube.

4.5 Inspecting

Inspect all product components after each cleaning and disin-
fection. Replace any broken, misshapen or seriously dis-
coloured parts.