Figure 11 Open the electrical enclosure
Figure 12 Replace a fuse
8.6 Replace the temperature controller
Electrocution hazard. Remove power from the instrument before doing maintenance or
service activities.
1. Remove power and stop the flow of sample and cooling water to the instrument. Refer to
Shutdown
on page 23.
2. Open the electrical enclosure. Refer to the illustrated steps in
3. Do the illustrated steps in
4. Close the electrical enclosure.
26 English
D A N G E R
Figure
13.
Figure 11
on page 26.