The 9525sc DCCP system removes CO
conductivity (CC) and degassed cation conductivity (DCC) and calculates pH. The system measures
a sample stream of high purity water. The system has three conductivity sensors and two controllers.
The top controller is connected to two sensors and shows the SC (channel 1) and CC (channel 2)
measurements. The bottom controller shows the DCC measurement. Refer to
Figure 2 9525sc DCCP system
1 Heater
2 Sample cooler
3 SC200 Controller (or Polymetron
9500 controller)
4 Polymetron 8315 conductivity sensor (3x)
5 Electrical enclosure
6 Flow cell (3x)
2
The power indicator light starts and stops with the heater.
8 English
and measures specific conductivity (SC), cation
2
7 Power switch
8 Power indicator light
9 Temperature controller
10 Sample flow meter
11 Sample flow valve
12 Resin column
Figure
2.
2