Hach 9525sc DCCP System Manual Del Usuario página 3

Ocultar thumbs Ver también para 9525sc DCCP System:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 114
Table of Contents
1 Specifications
on page 3
2 General information
3 Installation
on page 10
4 Startup
on page 17
5 User interface and navigation
6 Operation
on page 20
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Dimensions
Weight
Power requirements
Main supply voltage fluctuation ±10% of nominal voltage
Power consumption
Pollution degree
Installation category
Altitude
Operating temperature
Operating temperature
(recommended)
Storage temperature
8315 Conductivity sensor
Controller power requirements
Controller and instrument
rating
Sample water
on page 4
on page 18
Details
91 x 62 x 38 cm (36 x 24.5 x 15 inches)
Degas-only DCCP system : 27.7 kg (61 lb)
DCCP system: 50 kg (110 lb)
US: 110 to 120 VAC, 1 phase, 50/60 Hz
EU: 220 to 240 VAC, 1 phase, 50/60 Hz
1.6 kVA
2
II
2000 m (6562 ft) maximum
Without regenerative cooling option: 2 to 50 °C (36 to 122 °F), 0 to
85% relative humidity, non-condensing
With regenerative cooling option: 2 to 45 °C (36 to 113 °F), 0 to
85% relative humidity, non-condensing
23 to 27 °C (73 to 81 °F)
0 to 50 °C (32 to 122 °F)
Cell constant, k: 0.01 cm
Measurement range: 0.01 to 200 μS/cm
Sensor accuracy: less than ± 2%
Power to the sensor(s) is supplied by the controller.
100 to 240 VAC ±10%, 50/60 Hz
IP66/NEMA 4X
Flow rate: 100 to 150 mL/min; 6 to 9 L/hr (1.5 to 2.4 gal/hr)
Temperature: 25 °C ± 1 °C (77 °F ± 2 °F) recommended; 2 to
54 °C (36 to 129 °F) without regenerative cooling option or 2 to
45 °C (36 to 113 °F) with regenerative cooling option
Pressure: 6.9 bar (100 psig)
7 Advanced operation
8 Maintenance
on page 22
9 Troubleshooting
on page 28
10 Replacement parts and accessories
on page 29
-1
on page 21
English 3

Publicidad

Tabla de contenido
loading

Tabla de contenido