Cleaning Instructions; External Surfaces Of The Spirometer; Cleaning Accessories - CareFusion MicroRPM Manual De Funcionamiento

Medidor de presión respiratoria
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Cleaning Instructions

Disinfection of contaminated parts is only effective after careful
preliminary cleaning. CareFusion recommends the tested solution of
PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
The device must not be wiped with any aqueous solutions and must
not be exposed to solvents e.g. alcohol, chloride solutions, as there
are electronic components inside that will be permanently damaged.
CAUTION: Switch off the device before cleaning.

External Surfaces of the Spirometer

CAUTION: Do not attempt to wash or immerse the Respiratory
Pressure Meter in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged
The external housing of the pressure meter may be wiped with sterile
wipes or a damp cloth that has been immersed in a cold sterilizing
solution.

Cleaning accessories

The MicroRPM unit is protected from contamination by the Bacterial
Filter (36-FIL6050) during mouth pressures measurements.
The Rubber Flanged Mouthpiece (36-MTH6400), the Expiratory and
Inspiratory Pressure Valve Assembly (36-ASS1221, 36-ASS1222) and
the Nasal Probes (36-NPROBE01, 36-NPROBE02, 36-NPROBE03, 36-
NPROBE04), however, may be immersed in a cold sterilizing solution
such as PeraSafe (36-SSC5000A). Rinse thoroughly and allow to dry
before reassembly.
39

Publicidad

Tabla de contenido
loading

Tabla de contenido