smartcode and adherence score information

Your homecare provider may contact you to retrieve SmartCode and Adherence Score information. To display the
SmartCode, press the LEFT arrow key once. Use the UP arrow or DOWN arrow keys to change the SmartCode reporting
period. Press the LEFT arrow key again to display the Adherence Score.
The Quick View Menu shows a summary of information which is reported in the SmartCode data. Information in the Quick
View includes: Number of days used more than 4 hours (or can be set to 5 hours by clinician), AHI (Apnea/Hypopnea
Index), High Leak percent (percent of time at leak >95 L/m), and Daily Usage (average hours per day used). Each
parameter can be viewed in any of four time periods, 1 day, 7 days, 30 days and 90 days.
Compliance information can be viewed without turning the CPAP on. With the blower OFF (display shows OFF), press the
left or right arrow keys on the keypad to scroll through the information.

patieNt MessaGes

Your IntelliPAP will alert you of issues that may require some action on your part in order to make your therapy more
effective. You can press any key or button to clear the message from the display, but the IntelliPAP will operate normally
even if the message is not cleared. Two kinds of messages are displayed: Reminders and Notifications.

reminders

Reminders alert you when parts of your system need to be replaced. Some components of your system wear over time
and, if not replaced, may compromise your therapy. If you see a message on the display that is not listed below, you may
have the optional DeVilbiss SmartLink Module attached to your device (Fig. F), which provides additional messages. Refer
to the documentation that was provided with the SmartLink Module.
Clean Filter – The filter should be checked every 10 days for signs of dirt or wear so that it can be cleaned as needed.
This message helps to remind you to check it regularly.

Notifications

Notifications identify conditions in your IntelliPAP that may require action by you or your equipment provider in order to
maintain a high level of therapy. If you see a message on the display that is not listed below, you may have the optional
DeVilbiss SmartLink Module attached to your device (Fig. F), which provides additional messages. Refer to the
documentation that was provided with the SmartLink Module.
Delay Running xx Minutes Left – While a comfort delay session is active, this message is flashed on the display every 5
seconds to let you know how much time is left in the delay.
Mask Leak – This message means the IntelliPAP has detected a large amount of airflow during therapy for at least 10% of
the time during your previous use session. This notification is displayed when the device is turned on. If this message is
displayed, put on the mask and adjust the headgear to ensure the mask is properly fitted to your face. Follow the mask
manufacturer's directions for adjusting the mask and headgear for proper fitting.
Mask Off – This message appears due to a poor mask fit or a removed mask. Check for leak around the mask seal and
make adjustments as necessary according to the mask manufacturer's instructions. This message will be displayed until
the high airflow problem is corrected. If the high airflow condition persists for about 20 seconds, the device will
automatically turn off if Auto-OFF is enabled.
Device Fault – If a device fault message is displayed, refer to Troubleshooting for instructions.
A-DV57
This option controls the Auto-ON feature, which automatically starts the flow of air
after breathing once or twice into the mask. This is always enabled when Auto-OFF is
enabled. If the unlocked symbol is shown, you are able to adjust this setting. If the
locked symbol is shown, this setting can only be adjusted by your provider.
This option controls the mask fit check feature. The air flow to your mask is
constantly measured and if the amount of airflow exceeds 95 liters/minute for more
than 10% of the time used, a notification will be created. This notification will be
displayed the next time you use the IntelliPAP.
Low backlight Enable will keep the backlight dim during operation. Low backlight
Disabled will turn the backlight OFF during operation.
Pressing the up key exits the Enable menu.
eN - 9