Creased hose WM 24120 (grey) can be steam-sterilized using
equipment to EN 285. Temperature: 134 ˚C, minimum treatment
time 3 minutes. Follow EN 554/ ISO 11134 with regard to vali-
dation and monitoring.
Accessories
To disinfect the exhalation system, the mask, the VENTIpower, the VENTI-O
VENTI-O
plus, the bacteria filter and the VENTIclick, please see the section entitled
2
"Disinfection and sterilization" in the relevant operating instructions.

6.4 Change in patients

If the unit is being operated with a bacteria filter:
change the bacteria filter
or
sterilize the bacteria filter and change the particle filter located
inside it.
If the unit is being operated without a bacteria filter:
have the unit subjected to hygienic preparation by a specialist deal-
er on change of patients.